Orally Disintegrating Tablet Tester

PT-ODT

PT-ODT uses a dedicated single cup and stainless-steel basket to observe orally disintegrating tablets. A short 10 mm stroke and a seconds-based timer distinguish this apparatus from conventional basket testers.

Key features

  • Single-cup, single-basket arrangement for direct observation
  • Dedicated cup locates the test vessel securely
  • Adjustable timing with a 100-second startup default

Technical Information

Basket stations1
Basket tube length30 mm
Basket internal diameter13 mm
Screen aperture710 µm
Movement frequency30 cycles/min
Stroke10 ± 1 mm
Bath temperatureAmbient to 45 °C
Temperature control accuracy< ±0.3 °C
Timer range1–900 s
Timer accuracy±1 s

Standard Configuration

One dedicated disintegration basket and special test cup are described for the base instrument.

Applications

Laboratory disintegration observation for orally disintegrating tablets; confirm the method and endpoint procedure for the formulation.

Selection / Ordering Information

Confirm sample format, required method, accessories, power supply and destination-specific configuration in the quotation. Published operating functions do not establish certification or regulatory compliance.

Downloads

Download the detailed product datasheet for full specifications and configuration information.

Download PDF - English · PDF

PRECETTA-PT-ODT-en.pdf (PRECETTA-PT-ODT-en.pdf, 440 Kb) [Download]

PRECETTA-PT-ODT-es.pdf (PRECETTA-PT-ODT-es.pdf, 441 Kb) [Download]

PRECETTA-PT-ODT-pt.pdf (PRECETTA-PT-ODT-pt.pdf, 441 Kb) [Download]

PRECETTA-PT-ODT-fr.pdf (PRECETTA-PT-ODT-fr.pdf, 440 Kb) [Download]

PRECETTA-PT-ODT-ar.pdf (PRECETTA-PT-ODT-ar.pdf, 457 Kb) [Download]

You may be interested
  • Recently Viewed