T-PEN-001 — Measures the consistency of ointments and ophthalmic ointments by timed cone penetration with automatic release and non-contact displacement sensing.
Key features
- Automatic cone release and motorized lift
- Fine-adjustment sample platform with LED contact illumination
- Mean and RSD calculation with printed results
- Colour display; documented interface uses Chinese menus
Core specifications
| Penetration range | >65 mm |
|---|
| Resolution | 0.01 mm |
|---|
| Penetration time | 5–99 s |
|---|
| Standard cone mass | 102.5 ±0.05 g |
|---|
| Standard cone shaft mass | 47.5 ±0.05 g |
|---|
| Platform adjustment | ±25 mm; 0.1 mm |
|---|
Download the detailed product datasheet for full specifications and configuration information.
Standard Configuration
Instrument with automated cone mechanism, adjustable platform, LED light and result printing. Confirm which cone and container sets are included; the technical description does not give a consistent packing list.
Optional Accessories
Half-cone and quarter-cone assemblies with matching containers are orderable. Treat them as separately confirmed options until the quotation specifies otherwise.
Selection / Ordering Information
Specify sample type, cone size, container, penetration time and interface-language needs. Confirm the required pharmacopoeial method and electrical supply.
Downloads
Download the detailed product datasheet for full specifications and configuration information.
Download PDF — English · PDF