PC-Controlled Atomic Absorption Spectrophotometer

PT-AAS-002

Elemental analysis with PC control and flame atomization. Graphite furnace, hydride and nitrous-oxide configurations must be selected for the application; their availability does not imply inclusion in the base package.

Technical Information

ModelPT-AAS-002
Wavelength range190–900 nm
Wavelength accuracy±0.2 nm
Wavelength repeatability±0.1 nm
Absorbance range0–2.5 Abs
Spectral bandwidth0.1 / 0.2 / 0.4 / 1.0 / 1.2 / 2.0 nm
Optical resolution testMn 279.5 / 279.8 nm at 0.2 nm; valley/peak <30%
Baseline drift, static≤0.002 Abs/30 min (Cu)

Standard configuration

Main spectrophotometer with PC control functions and air-acetylene flame capability. Confirm the computer, lamps, lamp-position count and supplied accessories in the quotation.

Optional configuration

Nitrous-oxide flame function with 50 mm burner; graphite furnace and hydride-generation configurations. No graphite performance values are inferred from the flame data.

Downloads

Download the language-specific PDF for complete specifications, configuration and application guidance.

Download product PDF

Selection and ordering

Specify elements, matrix, target concentration and atomization method. Confirm lamp positions and complete gas/extraction/utility requirements before order.

You may be interested