Elemental analysis with PC control and flame atomization. Graphite furnace, hydride and nitrous-oxide configurations must be selected for the application; their availability does not imply inclusion in the base package.
Technical Information
| Model | PT-AAS-002 |
|---|
| Wavelength range | 190–900 nm |
|---|
| Wavelength accuracy | ±0.2 nm |
|---|
| Wavelength repeatability | ±0.1 nm |
|---|
| Absorbance range | 0–2.5 Abs |
|---|
| Spectral bandwidth | 0.1 / 0.2 / 0.4 / 1.0 / 1.2 / 2.0 nm |
|---|
| Optical resolution test | Mn 279.5 / 279.8 nm at 0.2 nm; valley/peak <30% |
|---|
| Baseline drift, static | ≤0.002 Abs/30 min (Cu) |
|---|
Standard configuration
Main spectrophotometer with PC control functions and air-acetylene flame capability. Confirm the computer, lamps, lamp-position count and supplied accessories in the quotation.
Optional configuration
Nitrous-oxide flame function with 50 mm burner; graphite furnace and hydride-generation configurations. No graphite performance values are inferred from the flame data.
Downloads
Download the language-specific PDF for complete specifications, configuration and application guidance.
Download product PDF
Selection and ordering
Specify elements, matrix, target concentration and atomization method. Confirm lamp positions and complete gas/extraction/utility requirements before order.