Automatic laboratory steam autoclaves with 145 L or 170 L chambers, pulse vacuum, vacuum drying and a separate steam generator.
- Automatic water, steam, cooling and drying sequence
- Supplied sample-temperature probe and built-in printer
- Touchscreen, user permissions and USB data export
Image shows PT-AUT-025. PT-AUT-026 has a taller chamber and cabinet.
Technical Information
| Process | Temperature range | Time range |
|---|
| Sterilization | 105–135 °C | 0–120 min |
| Drying | 60–135 °C | 0–240 min |
| Melting | 60–100 °C | 0–999 min |
| Holding | 40–60 °C | 0–1440 min |
| Parameter | Value |
|---|
| Design pressure | 0.38 MPa |
| Maximum working pressure | 0.24 MPa |
| Electrical supply | 380 V ±38 V / 50 Hz ±1 Hz |
| Model | Chamber volume (L) | Input power (kW) |
|---|
| PT-AUT-025 | 145 | 9.5 |
| PT-AUT-026 | 170 | 9.5 |
| Model | Chamber: diameter × height (mm) | External dimensions (mm) |
|---|
| PT-AUT-025 | Ø500 × 720 | 890 × 910 × 1200 |
| PT-AUT-026 | Ø500 × 850 | 890 × 910 × 1245 |
Standard Configuration
SUS304 stainless steel chamber and baskets, separate steam generator, pulse-vacuum and vacuum-drying functions, automatic water feed, condenser and fan cooling, sample-temperature probe and built-in printer. Touchscreen controller includes 40 sterilization programs and 10 melting/holding programs, USB data export and operator permissions. Electric lid opening has mechanical self-locking and electronic interlocking. Dual low-water detection and temperature/pressure protection are provided.
Optional Accessories
Water-storage system and waste-liquid collection tank are optional when site water supply or drainage is unavailable. Confirm connections, basket quantities and the final delivery list. Validation services are not stated as included.
Downloads
Download the complete datasheet for model specifications and configuration details.
Download PDF - English
Selection / Ordering Information
Specify model, load type, required process and site services. Confirm electrical supply, water, drainage, installation clearances and the cycle suitable for the intended load.