18 L and 24 L electrically heated steam sterilizers for laboratory use, with a listed maximum working temperature of 126 °C and low-water protection.
- SUS304 stainless steel outer housing
- Pressure relief and manual vent valve
- Low-water heater cutoff and end-of-cycle alarm
Representative exterior view. The two models differ in height and capacity; lid and controller style are specific to this series.
Technical Information
| Parameter | Value |
|---|
| Maximum working temperature | 126 °C |
| Temperature resolution | 0.1 °C |
| Maximum working pressure | 0.142 MPa |
| Design pressure | 0.165 MPa |
| Electrical supply | 220 V |
| Input power (W) | 2000 W |
| Timer range | 0–999 min |
| Model | Chamber volume (L) | Inner container: diameter × height × quantity (mm) |
|---|
| PT-AUT-036 | 18 | Ø280 × 230 × 1 |
| PT-AUT-037 | 24 | Ø280 × 330 × 1 |
Standard Configuration
SUS304 outer housing, electric heater, pressure gauge, pressure relief valve, manual vent valve, low-water safety float, heater cutoff with audible alarm, silicone seal and end-of-cycle alarm with automatic heating shutdown. Cooling is natural. Digital display for set temperature, actual temperature and cycle time.
Optional Accessories
Confirm containers, accessories and the required laboratory cycle in the quotation. No printer, software or validation service is listed as standard.
Downloads
Download the complete datasheet for model specifications and configuration details.
Download PDF - English
Selection / Ordering Information
Specify the model, load or sample type, operating temperature and required basket configuration. Confirm electrical supply, accessories and final delivery before ordering.