Programmable forced-air ovens for laboratory drying and thermal processing, with 65–452 L chambers and motorized airflow adjustment.
- 10-segment, 100-cycle programming
- Six fan speeds and motorized air damper
- Dual overtemperature alarms and RS-485 interface
Image shows PT-OVN-037. Other models differ in dimensions.
Technical Information
| Parameter | Value |
|---|
| Temperature range | Ambient +5 to 300 °C |
| Temperature resolution | 0.1 °C |
| Timer range | 0–9999 min/h |
| Load per shelf | 25 kg |
| Electrical supply | 220 V ±10 V / 50–60 Hz |
| Delayed-start range | 0–9999 min |
| Temperature uniformity (Ambient +5 to 105 °C) | ±1% |
| Temperature uniformity (105–300 °C) | ±2% |
| Model | Chamber volume (L) | Input power (W) | Shelves: supplied / maximum |
|---|
| PT-OVN-037 | 65 | 1560 | 2/5 |
| PT-OVN-038 | 130 | 2160 | 2/7 |
| PT-OVN-039 | 210 | 2760 | 3/7 |
| PT-OVN-040 | 284 | 2760 | 3/9 |
| PT-OVN-041 | 452 | 3500 | 4/13 |
| Model | Temperature fluctuation (150 °C) | Time to 150 °C (min) |
|---|
| PT-OVN-037 | ±0.3 °C | 17 |
| PT-OVN-038 | ±0.3 °C | 20 |
| PT-OVN-039 | ±0.5 °C | 20 |
| PT-OVN-040 | ±0.5 °C | 20 |
| PT-OVN-041 | ±0.5 °C | 22 |
| Model | Chamber dimensions (mm) | External dimensions (mm) |
|---|
| PT-OVN-037 | 410 × 350 × 450 | 555 × 566 × 722 |
| PT-OVN-038 | 570 × 400 × 570 | 715 × 616 × 842 |
| PT-OVN-039 | 630 × 530 × 630 | 775 × 746 × 902 |
| PT-OVN-040 | 720 × 530 × 745 | 865 × 746 × 1019 |
| PT-OVN-041 | 600 × 580 × 1300 | 774 × 845 × 1704 |
Standard Configuration
Programmable controller with backlit touch keys, six-speed fan, motorized air damper, preheating chamber, slide-out shelves, dual-stage door lock and one RS-485 interface. Overtemperature protection uses the temperature sensor system and an independent temperature switch.
Optional Accessories
Inlet and exhaust air filters, extra shelves, printer, monitoring software and GPRS alarm messaging are optional. Confirm compatibility and scope in the quotation. No software compliance or validation service is included by default.
Downloads
Download the complete datasheet for model specifications and configuration details.
Download PDF - English
Selection / Ordering Information
Specify the model, sample or application, working temperature and required shelf arrangement. Confirm electrical supply and the final configuration before ordering.