Precision Programmable Laboratory Oven Series

PT-OVN-037

Programmable forced-air ovens for laboratory drying and thermal processing, with 65–452 L chambers and motorized airflow adjustment.

  • 10-segment, 100-cycle programming
  • Six fan speeds and motorized air damper
  • Dual overtemperature alarms and RS-485 interface

Image shows PT-OVN-037. Other models differ in dimensions.

Technical Information

ParameterValue
Temperature rangeAmbient +5 to 300 °C
Temperature resolution0.1 °C
Timer range0–9999 min/h
Load per shelf25 kg
Electrical supply220 V ±10 V / 50–60 Hz
Delayed-start range0–9999 min
Temperature uniformity (Ambient +5 to 105 °C)±1%
Temperature uniformity (105–300 °C)±2%
ModelChamber volume (L)Input power (W)Shelves: supplied / maximum
PT-OVN-0376515602/5
PT-OVN-03813021602/7
PT-OVN-03921027603/7
PT-OVN-04028427603/9
PT-OVN-04145235004/13
ModelTemperature fluctuation (150 °C)Time to 150 °C (min)
PT-OVN-037±0.3 °C17
PT-OVN-038±0.3 °C20
PT-OVN-039±0.5 °C20
PT-OVN-040±0.5 °C20
PT-OVN-041±0.5 °C22
ModelChamber dimensions (mm)External dimensions (mm)
PT-OVN-037410 × 350 × 450555 × 566 × 722
PT-OVN-038570 × 400 × 570715 × 616 × 842
PT-OVN-039630 × 530 × 630775 × 746 × 902
PT-OVN-040720 × 530 × 745865 × 746 × 1019
PT-OVN-041600 × 580 × 1300774 × 845 × 1704

Standard Configuration

Programmable controller with backlit touch keys, six-speed fan, motorized air damper, preheating chamber, slide-out shelves, dual-stage door lock and one RS-485 interface. Overtemperature protection uses the temperature sensor system and an independent temperature switch.

Optional Accessories

Inlet and exhaust air filters, extra shelves, printer, monitoring software and GPRS alarm messaging are optional. Confirm compatibility and scope in the quotation. No software compliance or validation service is included by default.