Digital Multi-Compartment Pharmaceutical Stability Chambers

PT-ENV-108

Run different stability conditions in independently controlled compartments. Select the compartment layout and light-exposure capability that match your study.

Technical Information

ModelPT-ENV-108
Temperature range (°C)15-65
Temperature fluctuation (°C)±0.5
Temperature deviation (°C)±2
Humidity range (% RH)15-95
Humidity deviation (% RH)±5
Power supply220 V / 50 Hz
Operating ambient conditions5-35 °C; ≤85% RH

Configuration A

Illuminance (lx)Not fitted
Illuminance deviation (lx)Not fitted
Compartment volume (L)2 × 250 = 500
Installed power (W)3000
Internal dimensions (mm)2 × (500×500×1000)
External dimensions (mm)1050×1340×1860
Packed dimensions (mm)1100×1440×2100
Net mass (kg)390

Configuration B

Illuminance (lx)10-10000
Illuminance deviation (lx)±500
Compartment volume (L)2 × 250 = 500
Installed power (W)3000
Internal dimensions (mm)2 × (500×500×1000)
External dimensions (mm)1050×1340×1860
Packed dimensions (mm)1100×1440×2100
Net mass (kg)390

Configuration C

Illuminance (lx)10-10000
Illuminance deviation (lx)±500
Compartment volume (L)Confirm with quotation
Installed power (W)4300
Internal dimensions (mm)2 × (500×500×500) + 500×500×1000
External dimensions (mm)1050×1340×1860
Packed dimensions (mm)1100×1440×2100
Net mass (kg)400

Configuration D

Illuminance (lx)10-10000
Illuminance deviation (lx)±500
Compartment volume (L)2 × 125 + 2 × 250 = 750
Installed power (W)4800
Internal dimensions (mm)2 × (500×500×1000) + 2 × (500×500×500)
External dimensions (mm)1050×2110×1860
Packed dimensions (mm)1100×2210×2100
Net mass (kg)580

Performance values are specified for the selected configuration. Do not combine temperature or humidity limits from different configurations. Supply is listed as 220 V / 50 Hz; other electrical requirements need confirmation.

Equipment & Construction

Forced-air circulation, digital temperature control, a stainless-steel interior and overtemperature protection. The listed configuration determines chamber dimensions and shelf quantity. Humidity control uses steam humidification and an automatic circulating water system.

Configuration Options

Confirm recording, monitoring, qualification documents and any additional accessories in the quotation. These are not a blanket certification or a promise of standard supply.

Downloads

Technical data and configuration details for this instrument series.

Download technical data sheet (PDF)

Ordering Information

Specify the series model, configuration letter, sample load, setpoints, installation space and destination power requirements. All dimensions are depth × width × height. Confirm the final configuration and applicable test protocol before ordering.