T-INH-001 — A mobile laboratory test station with vacuum and flow measurement for inhalation-product sampling. The documented system combines a glass particle/droplet-distribution apparatus with a selectable dose-uniformity sampling configuration.
Key features
- Mobile cabinet designed as a test workbench
- Digital LCD flow display
- Continuous manual flow adjustment and on/off control
- Selectable six- or ten-tube dose-uniformity sampling
Technical Information
| Pump no-load flow | >110 SLPM |
|---|
| Flow measurement range | 10–100 SLPM |
|---|
| Flow measurement accuracy | ±5% |
|---|
| Display resolution | 0.001 SLPM |
|---|
| Electrical supply | 220 V / 50 Hz / 2.5 A |
|---|
| Dimensions | 650 ×450 ×1200 mm |
|---|
| Approximate mass | 45 kg |
|---|
| Pump operating noise | <80 dB(A) |
|---|
Download the detailed product datasheet for full specifications and configuration information.
Standard Configuration
Mobile main cabinet and one glass particle/droplet-distribution apparatus are documented. Confirm the vacuum pump, sampling tubes, adapters and complete supplied configuration.
Optional Accessories
Dose-uniformity sampling is selectable with six or ten tubes. Confirm the chosen set and accessory compatibility; do not assume both sets are supplied.
Selection / Ordering Information
Specify inhaler type, sampling method and required flow. No specific NGI/ACI stages or aerodynamic cut-off diameters are established by the available information. Display resolution is not measurement accuracy. Confirm electrical supply and method configuration.
Downloads
Download the detailed product datasheet for full specifications and configuration information.
Download PDF — English · PDF