Moves samples between programmed media rows for release profiling. Adjustable reciprocation, immersion and draining times support method development with selectable single- or double-row configurations.
Key features
- Programmable media transfer
- Selectable single- and double-row operation
- Adjustable immersion and draining times
- Two documented reciprocation strokes
Technical Information
| Bath heating range | ambient–45 °C |
|---|
| Temperature resolution | 0.1 °C |
|---|
| Temperature control error | ≤±0.2 °C |
|---|
| Reciprocation frequency | 1–50 min⁻¹ |
|---|
| Frequency error | ±5% |
|---|
| Cylinder-method stroke | 10 ±0.1 cm |
|---|
| Holder-method stroke | 2 cm |
|---|
| Heater power | 1500 W |
|---|
| Electrical supply | 220 V / 50 Hz / 10 A |
|---|
| Operating temperature | 5–40 °C |
|---|
| Relative humidity | ≤80% |
|---|
| Operating noise | ≤60 dB |
|---|
| Vessel configurations | 50–300 mL: 7 / 14; 500–1000 mL: 6 |
|---|
Download the detailed product datasheet for full specifications and configuration information.
Standard Configuration
Instrument in the documented configuration; confirm the complete delivery list, software licence and computer inclusion in the quotation.
Optional Accessories
Confirm method accessories, compatible sampling equipment and qualification documents separately. Availability and cost depend on the selected configuration.
Selection / Ordering Information
Confirm method edition, sample format, vessel configuration and documentation needs. Method references and software audit features do not establish certification or validated compliance. Heating specifications do not establish active cooling.
Downloads
Download the detailed product datasheet for full specifications and configuration information.
Download PDF — English · PDF