Biological safety cabinets with LCD airflow monitoring, HEPA filtration, stainless steel work areas and model-specific cabinet dimensions.
- Automatic airflow adjustment and LCD display
- UV/lighting interlock and front-window alarm
- Removable SUS304 worktop and drain
Representative cabinet. Work-area width, lighting and exhaust arrangement depend on the selected model; confirm the final installed configuration.
Technical Information
| Parameter | Value |
|---|
| Average inflow velocity | 0.55 m/s |
| Downflow control precision | ±0.015 m/s |
| Operating temperature | 10–30 °C |
| Operating RH at 25 °C | ≤70% |
| Average illuminance | ≥650 lx |
| Electrical supply | AC 220 V ±10 V / 50 / 60 Hz |
| Differential-pressure display | 0–500 Pa |
| Model | Average downflow velocity | Power including outlet (kW) |
|---|
| PT-BSC-004 | 0.35 m/s | 1.8 |
| PT-BSC-005 | 0.33 m/s | 2 |
Standard Configuration
LCD monitoring, automatic airflow adjustment, filter fault alarms, UV/lighting interlock and front-window position alarm. Standard HEPA filtration, observation lighting and UV lamp, SUS304 work area with removable worktop, splash-protected AC outlet, reserved water/gas connections, collection trough and drain valve. An exhaust hood and duct connection are provided for a central exhaust system or an optional external exhaust fan. The supply fan stops when the front window is closed; the base has casters and leveling adjustment.
Optional Accessories
Confirm filter grade, supplied service fittings, installation and exhaust requirements, commissioning and qualification documentation in the quotation. No third-party certification or protection test result is asserted by this datasheet. The external exhaust fan is optional; confirm the building exhaust arrangement.
Downloads
Download the complete datasheet for model specifications and configuration details.
Download PDF - English
Selection / Ordering Information
Specify the PRECETTA model, work-area width, intended process, site utilities and exhaust arrangement. Confirm filtration, commissioning and required qualification documents before ordering.