T-SOF-001 — Determines the softening point of medicinal plasters in the 40–80 °C range using controlled water-bath heating and automatic optical detection.
Key features
- Automatic temperature control throughout the test
- Automatic steel-ball placement after equilibration
- Fibre-optic endpoint detection
- Heating without stirring to avoid disturbing samples
Core specifications
| Sample softening-point range | 40–80 °C |
|---|
| Display resolution | 0.1 °C |
|---|
| Temperature error at 30–50 °C | ±0.2 °C |
|---|
| Temperature error at 50–80 °C | ±0.3 °C |
|---|
| Bath heating rate | 1.0–1.5 °C/min |
|---|
| Bath equilibrium temperature | 37 ±1 °C |
|---|
Download the detailed product datasheet for full specifications and configuration information.
Standard Configuration
Instrument with controlled bath heating, ball-placement mechanism and optical detection. Confirm consumables and the complete supplied fixture list.
Optional Accessories
No optional accessory list is documented. Request compatible spare test parts and qualification-document availability in the quotation.
Selection / Ordering Information
Specify the plaster formulation, expected softening range and applicable method edition. Confirm electrical supply and required test fixtures before ordering.