50–100 L vertical steam sterilizers with a supplied drying function, separate sterilization and drying timers and model-specific heater ratings.
- Handwheel lid and pressure safety interlock
- Low-water, overtemperature and overpressure protection
- Digital temperature and timer controls
Representative exterior view. Cabinet proportions, chamber dimensions and supplied basket arrangement vary by model.
Technical Information
| Parameter | Value |
|---|
| Maximum working temperature | 135 °C |
| Temperature resolution | 1 °C |
| Maximum working pressure | 0.22 MPa |
| Design pressure | 0.25 MPa |
| Electrical supply | 220 V |
| Timer range | 4–120 min |
| Drying timer | 30–240 min |
| Model | Chamber volume (L) | Sterilization heater power (W) | Drying heater power (W) |
|---|
| PT-AUT-009 | 50 | 3200 | 800 |
| PT-AUT-010 | 75 | 3500 | 1000 |
| PT-AUT-011 | 100 | 3500 | 1000 |
Standard Configuration
SUS304 construction, handwheel lid, microcomputer control, self-expanding seal, automatic cold-air discharge, low-water control, overtemperature protection, pressure relief, pressure safety interlock and end-of-cycle audible alarm. Internal exhaust circulation and drying. Modes: sterilization with natural cooling; sterilization with automatic exhaust; or sterilization, exhaust, drainage and drying.
Optional Accessories
Confirm additional containers and accessories in the quotation. Printer, software and validation services are not listed as standard. Optional melting/holding function: melting 60–98 °C and holding 40–60 °C, each with 0–999 minute timing. Confirm the selected option.