35–120 L vacuum drying ovens with independently controlled heated shelves, ambient +5 to 250 °C settings, USB data export and an event audit trail.
- Independent heating control for each shelf
- USB / RS-485 interfaces and PDF / CSV data export
- User permissions, event records and audit trail
Representative exterior. Select the model, cabinet quantity and holder or shelf configuration from the specifications; the photograph does not define the complete delivery.
Technical Information
| Parameter | Value |
|---|
| Temperature range | Ambient +5 to 250 °C |
| Temperature control precision | ±0.1 °C |
| Temperature fluctuation at 100 °C | ±0.2 °C |
| Timer range | 0–9999 min / h |
| Vacuum display range | 0.0–101 kPa |
| Vacuum display precision | 0.1 kPa |
| Ultimate vacuum | 133 Pa |
| Listed leakage rate | 1.0 kPa/h |
| Inert-gas backfill pressure | ≤101 kPa |
| Gas connector outside diameter | 12 mm |
| Supplied shelves | 2 |
| Load per shelf | 10 kg |
| Electrical supply | AC 220 V ±10% / 50 Hz |
| Model | Chamber volume (L) | Input power (W) | Temperature uniformity at 100 °C |
|---|
| PT-VDO-044 | 35 | 900 | ±2 °C |
| PT-VDO-045 | 65 | 1100 | ±2 °C |
| PT-VDO-046 | 120 | 1300 | ±3 °C |
Standard Configuration
Stainless steel chamber, two aluminium-alloy heated shelves with independent temperature control and mechanical temperature limits, observation door, touchscreen controller, parameter memory, USB and RS-485 interfaces. Standard data management includes PDF/CSV export, assigned operator permissions and timestamped event audit records. Protection covers leakage current, overtemperature, vacuum timeout and sensor alarms. No regulatory certification or compliance claim is implied.
Optional Accessories
Vacuum pump and pump sound enclosure are optional. Remote monitoring or messaging services require a separately confirmed option and availability; they are not standard. Confirm pump compatibility, tubing, gas-pressure reference and final configuration.
Downloads
Download the complete datasheet for model specifications and configuration details.
Download PDF - English
Selection / Ordering Information
Specify the PRECETTA model, sample, operating conditions and required configuration. Confirm electrical supply, accessories and the final delivery list before ordering.