Floor-Standing Thermal Aging Oven - 300 °C

PT-ENV-052

Programmable chambers for thermal aging and temperature exposure, with ambient +5 to 300 °C settings, six fan speeds and model-specific chamber sizes.

  • Six selectable fan speeds
  • 10-segment programs with 100 cycles
  • Temperature-sensor and independent thermostat alarms

Representative exterior; dimensions and shelf configuration follow the selected model.

Technical Information

ParameterValue
Temperature rangeAmbient +5 to 300 °C
Program segments / cycles10 / 100
Temperature resolution0.1 °C
Timer range0–9999 min/h
Electrical supply220 V ±10 V / 50 / 60 Hz
Listed uniformity: ambient +5 to 105 °C±1%
Listed uniformity: 105 to 300 °C±2%
Listed shelf load25 kg
ModelChamber volume (L)Input power (W)
PT-ENV-0524523500
ModelTemperature fluctuation at 150 °C
PT-ENV-052±0.5 °C

Standard Configuration

LCD temperature/status display, six fan speeds, 10-segment/100-cycle programming, delayed start, sensor-based overtemperature alarm and an independent thermostat. Preheating chamber, sliding shelves and one RS-485 interface are included. Motor-adjusted air damper, rounded chamber corners, reinforced shelves and a two-stage rotary door latch.

Optional Accessories

UV lamp and inlet/outlet air filters are optional. An external printer or remote alarm/monitoring/control service requires a separately confirmed selection and availability; these are not supplied together as standard. Confirm filter testing, software, validation, accessories and final delivery in the quotation.

Downloads

Download the complete datasheet for model specifications and configuration details.

Download PDF - English

Selection / Ordering Information

Specify the PRECETTA model, sample, working temperature, shelf load and required options. Confirm mains supply, installation, thermal test conditions and delivery before ordering.