85 °C Programmable Laboratory Incubator Series

PT-INC-038

Programmable fan-circulated incubators with 40–1001 L chambers, ambient +5 to 85 °C control and model-specific thermal tolerances.

  • LCD controller: 10 segments and 100 cycles
  • Six fan-speed settings
  • Standard RS-485 interface

Representative exterior view. Cabinet dimensions, observation-door design and shelf quantities vary by model; consult the specification table.

Technical Information

ParameterValue
Temperature rangeAmbient +5 to 85 °C
Temperature resolution0.1 °C
Electrical supplyAC 220 V ±10% / 50 Hz ±2%
Timer range0–9999 min/h
ModelChamber volume (L)Input power (W)Shelves: supplied / maximum
PT-INC-038403002/4
PT-INC-039774502/5
PT-INC-0401424503/7
PT-INC-0412347503/10
PT-INC-04245210004/18
PT-INC-04365515004/18
PT-INC-044100122504/18
ModelTemperature fluctuationTemperature uniformity
PT-INC-038±0.5 °C (37 °C)±1 °C (37 °C)
PT-INC-039±0.5 °C (37 °C)±1 °C (37 °C)
PT-INC-040±0.5 °C (37 °C)±1 °C (37 °C)
PT-INC-041±0.5 °C (37 °C)±1 °C (37 °C)
PT-INC-042±0.8 °C (37 °C)±2%
PT-INC-043±0.8 °C (37 °C)±2%
PT-INC-044±0.8 °C (37 °C)±2%

Standard Configuration

Stainless steel chamber, adjustable shelves, LCD temperature/timer controller, 10-segment/100-cycle programming, six fan speeds, RS-485, overtemperature alarm and power-restoration memory. The observation window is documented only for PT-INC-038 to PT-INC-041; glazing on the three larger models requires confirmation.

Optional Accessories

Confirm additional shelves and any required accessories in the quotation. Lighting, UV lamps and software are not listed as supplied.

Downloads

Download the complete datasheet for model specifications and configuration details.

Download PDF - English

Selection / Ordering Information

Specify the model, sample or application, working temperature and required shelf arrangement. Confirm electrical supply and the final configuration before ordering.

You may be interested