Combine temperature, humidity and light exposure in one chamber for pharmaceutical stability investigations. Light and UV capabilities depend on the selected configuration.
Technical Information
| Model | PT-ENV-105 |
|---|
| Temperature range (°C) | 15-65 |
|---|
| Temperature fluctuation (°C) | ±0.5 |
|---|
| Temperature deviation (°C) | ±2 |
|---|
| Humidity range (% RH) | 15-95 |
|---|
| Humidity deviation (% RH) | ±5 |
|---|
| Illuminance (lx) | 10-10000 |
|---|
| Illuminance deviation (lx) | ±500 |
|---|
| Power supply | 220 V / 50 Hz |
|---|
| Operating ambient conditions | 5-35 °C; ≤85% RH |
|---|
Configuration A
| Illuminated shelves | 1 |
|---|
| Installed power (W) | 2000 |
|---|
| Internal dimensions (mm) | 440×500×570 |
|---|
| External dimensions (mm) | 950×690×1330 |
|---|
| Packed dimensions (mm) | 1040×800×1510 |
|---|
| Net mass (kg) | 160 |
|---|
| Shelves | 2 |
|---|
Configuration B
| Illuminated shelves | 1 |
|---|
| Installed power (W) | 2000 |
|---|
| Internal dimensions (mm) | 440×500×920 |
|---|
| External dimensions (mm) | 950×690×1670 |
|---|
| Packed dimensions (mm) | 1040×800×1870 |
|---|
| Net mass (kg) | 200 |
|---|
| Shelves | 4 |
|---|
Configuration C
| Illuminated shelves | 1 |
|---|
| Installed power (W) | 2500 |
|---|
| Internal dimensions (mm) | 600×600×1000 |
|---|
| External dimensions (mm) | 1110×790×1760 |
|---|
| Packed dimensions (mm) | 1200×880×1950 |
|---|
| Net mass (kg) | 230 |
|---|
| Shelves | 4 |
|---|
Configuration D
| Illuminated shelves | 1 |
|---|
| Installed power (W) | 2500 |
|---|
| Internal dimensions (mm) | 600×800×1000 |
|---|
| External dimensions (mm) | 1110×990×1760 |
|---|
| Packed dimensions (mm) | 1200×1090×1950 |
|---|
| Net mass (kg) | 270 |
|---|
| Shelves | 4 |
|---|
Performance values are specified for the selected configuration. Do not combine temperature or humidity limits from different configurations. Supply is listed as 220 V / 50 Hz; other electrical requirements need confirmation.
Equipment & Construction
Forced-air circulation, digital temperature control, a stainless-steel interior and overtemperature protection. The listed configuration determines chamber dimensions and shelf quantity. Humidity control uses steam humidification and an automatic circulating water system.
Configuration Options
Confirm recording, monitoring, qualification documents and any additional accessories in the quotation. These are not a blanket certification or a promise of standard supply.
Ordering Information
Specify the series model, configuration letter, sample load, setpoints, installation space and destination power requirements. All dimensions are depth × width × height. Confirm the final configuration and applicable test protocol before ordering.