Temperature, Humidity and Visible-Light Stability Chambers

PT-ENV-105

Combine temperature, humidity and light exposure in one chamber for pharmaceutical stability investigations. Light and UV capabilities depend on the selected configuration.

Technical Information

ModelPT-ENV-105
Temperature range (°C)15-65
Temperature fluctuation (°C)±0.5
Temperature deviation (°C)±2
Humidity range (% RH)15-95
Humidity deviation (% RH)±5
Illuminance (lx)10-10000
Illuminance deviation (lx)±500
Power supply220 V / 50 Hz
Operating ambient conditions5-35 °C; ≤85% RH

Configuration A

Illuminated shelves1
Installed power (W)2000
Internal dimensions (mm)440×500×570
External dimensions (mm)950×690×1330
Packed dimensions (mm)1040×800×1510
Net mass (kg)160
Shelves2

Configuration B

Illuminated shelves1
Installed power (W)2000
Internal dimensions (mm)440×500×920
External dimensions (mm)950×690×1670
Packed dimensions (mm)1040×800×1870
Net mass (kg)200
Shelves4

Configuration C

Illuminated shelves1
Installed power (W)2500
Internal dimensions (mm)600×600×1000
External dimensions (mm)1110×790×1760
Packed dimensions (mm)1200×880×1950
Net mass (kg)230
Shelves4

Configuration D

Illuminated shelves1
Installed power (W)2500
Internal dimensions (mm)600×800×1000
External dimensions (mm)1110×990×1760
Packed dimensions (mm)1200×1090×1950
Net mass (kg)270
Shelves4

Performance values are specified for the selected configuration. Do not combine temperature or humidity limits from different configurations. Supply is listed as 220 V / 50 Hz; other electrical requirements need confirmation.

Equipment & Construction

Forced-air circulation, digital temperature control, a stainless-steel interior and overtemperature protection. The listed configuration determines chamber dimensions and shelf quantity. Humidity control uses steam humidification and an automatic circulating water system.

Configuration Options

Confirm recording, monitoring, qualification documents and any additional accessories in the quotation. These are not a blanket certification or a promise of standard supply.