Double-Door Wide-Range Photostability Chambers

PT-ENV-030

Humidity-controlled photostability chambers with continuously adjustable 0–10000 lx shelf lighting, a 7-inch touchscreen and an audit trail.

  • 7-inch touchscreen with audit trail
  • 30 program segments and 99 cycles
  • Standard embedded printer; USB / RS-232 / RS-485

Representative exterior. Dimensions, shelves and internal modules follow the selected model and configuration.

Technical Information

Humidity / illumination stateTemperature range
Humidity off / light off0–85 °C
Humidity off / light on10–85 °C
Humidity on / light off10–60 °C
Humidity on / light on15–60 °C
ParameterValue
Temperature resolution0.1 °C
Temperature fluctuation at 25 °C±0.3 °C
Temperature uniformity at 25 °C±1 °C
Program segments / cycles30 / 99
Time per segment1–9999 h
Listed humidity range20–98%
Listed humidity fluctuation±3%
Illuminance: continuously adjustable0–10000 lx
Listed illuminance deviation±500 lx
ModelInput power (W)Shelves: supplied / maximum
PT-ENV-03034504 / 12
PT-ENV-03135504 / 12
PT-ENV-03236504 / 12
ModelIncluded light modules (each includes one source)Additional follower sources includedMaximum lighting positions, including modules
PT-ENV-030114
PT-ENV-031114
PT-ENV-032114

Standard Configuration

7-inch touchscreen with USB export, event audit trail and user permissions. Includes variable-speed refrigeration, antifog glass inner door, mechanical lock, embedded printer, USB, RS-232 and RS-485, two side validation ports (30 mm inside diameter) and condensate recovery. Adjustable circulation airflow and multilevel passwords are included.Humidity control and one shelf-lighting module including one light source are standard. Additional follower sources and maximum lighting positions vary by model; refer to the configuration table. Humidification uses surface evaporation outside the working chamber.

Optional Accessories

Near-UV module is optional: listed 0–2 W/m², 320–400 nm; it is not a germicidal-UV claim. Additional light sources follow the control module and have no independent display or printing. Extra shelves, software, remote messaging and qualification documents require quotation confirmation. No regulatory certification is assured. Confirm the selected modules and combined operating conditions. Power for simultaneous light and UV is not specified.

Downloads

Download the complete datasheet for model specifications and configuration details.

Download PDF - English

Selection / Ordering Information

Specify the PRECETTA model, temperature, humidity, illumination/UV needs and required modules. Confirm combined operating conditions, utilities, source quantities, installation and delivery before ordering.

You may be interested