Technical Information

ModelPT-ENV-103
Temperature range (°C)15-65
Temperature fluctuation (°C)±0.5
Temperature deviation (°C)±2
Humidity range (% RH)15-95
Humidity deviation (% RH)±5
Power supply220 V / 50 Hz
Operating ambient conditions5-35 °C; ≤85% RH

Configuration A

Compressor sets2
Installed power (W)1900
Internal dimensions (mm)440×500×570
External dimensions (mm)950×690×1330
Packed dimensions (mm)1040×800×1510
Net mass (kg)160
Shelves2

Configuration B

Compressor sets2
Installed power (W)1900
Internal dimensions (mm)440×500×920
External dimensions (mm)950×690×1670
Packed dimensions (mm)1040×800×1870
Net mass (kg)200
Shelves4

Configuration C

Compressor sets2
Installed power (W)2400
Internal dimensions (mm)600×600×1000
External dimensions (mm)1110×790×1760
Packed dimensions (mm)1200×880×1950
Net mass (kg)230
Shelves4

Configuration D

Compressor sets2
Installed power (W)2400
Internal dimensions (mm)600×800×1000
External dimensions (mm)1110×990×1760
Packed dimensions (mm)1200×1090×1950
Net mass (kg)270
Shelves4

Configuration E

Compressor sets3
Installed power (W)2800
Internal dimensions (mm)1000×1000×1000
External dimensions (mm)1510×1190×1760
Packed dimensions (mm)1600×1290×1950
Net mass (kg)380
Shelves4

Performance values are specified for the selected configuration. Do not combine temperature or humidity limits from different configurations. Supply is listed as 220 V / 50 Hz; other electrical requirements need confirmation.

Equipment & Construction

Forced-air circulation, digital temperature control, a stainless-steel interior and overtemperature protection. The listed configuration determines chamber dimensions and shelf quantity. Humidity control uses steam humidification and an automatic circulating water system.

Configuration Options

Confirm recording, monitoring, qualification documents and any additional accessories in the quotation. These are not a blanket certification or a promise of standard supply.

Downloads

Technical data and configuration details for this instrument series.

Download technical data sheet (PDF)

Ordering Information

Specify the series model, configuration letter, sample load, setpoints, installation space and destination power requirements. All dimensions are depth × width × height. Confirm the final configuration and applicable test protocol before ordering.

You may be interested
  • Recently Viewed