Combines eight dissolution positions with synchronized sample introduction and individual vessel-temperature monitoring. A height-positioned sampling rack can connect to a compatible external autosampler.
Key features
- Synchronized automatic sample introduction
- Individual vessel-temperature monitoring
- Automatic sampling-rack height positioning
- Three access levels with documented audit records
Technical Information
| Working shafts | 8 |
|---|
| Bath vessels | 8 |
|---|
| Speed range | 5–250 rpm |
|---|
| Speed error | ±1% |
|---|
| Temperature control error | ±0.2 °C |
|---|
| Shaft / vessel axis deviation | ≤1.0 mm |
|---|
| Paddle wobble | ≤0.5 mm |
|---|
| Basket wobble | ≤1.0 mm |
|---|
| Speed setting increment | 1 rpm |
|---|
| Temperature display resolution | 0.1 °C |
|---|
| Bath heating range | ambient–45 °C |
|---|
| Electrical supply | 220 V / 50 Hz / 10 A |
|---|
Download the detailed product datasheet for full specifications and configuration information.
Standard Configuration
Instrument in the documented vessel/shaft configuration. Confirm the delivery list, baskets, paddles, glassware and software licence; method accessories and external sampling equipment are not assumed included.
Optional Accessories
Confirm compatible sampling equipment, printer, vessel certificates and qualification documents separately. Availability, interface and cost depend on the selected configuration.
Selection / Ordering Information
Bath heating extends from ambient temperature to 45 °C; a 5 °C label does not establish active cooling. Operating environment 5–37 °C, RH ≤80%. Confirm method, volume, accessory compatibility and software version. Documented method references do not establish certification or validated regulatory compliance.
Downloads
Download the detailed product datasheet for full specifications and configuration information.
Download PDF — English · PDF