Counts particles in liquids with large-volume sampling for infusion-device contamination and filtration testing. Eight configurable channels also support injectable solution and packaging-material investigations.
Key features
- Laser light-obscuration measurement
- Configurable size thresholds for method selection
- Stored results and printed reports
- Confirm the required method and delivered software configuration
Technical Information
| Particle size range | 2–400 µm |
|---|
| Sampling volume | 0.5–1000 mL |
|---|
| Concentration range | 0–10000 particles/mL |
|---|
| Sampling volume error | <±1% |
|---|
| Accuracy against reference value | ±10% |
|---|
| Repeatability at >1000 particles/mL, 5 mL sample | RSD <2% |
|---|
| Channels | 8 |
|---|
| Sampling rate | 15–150 mL/min |
|---|
| Electrical supply | 100–264 V; 50 Hz ±1%; <80 W |
|---|
Download the detailed product datasheet for full specifications and configuration information.
Standard Configuration
Instrument with the documented functions. Confirm all supplied holders, glassware, cables, reagents and consumables in the quotation; unspecified items are not assumed included.
Optional Accessories
No complete separate accessory list is available. Request compatible consumables, calibration items and qualification documents separately, with availability and price confirmed before ordering.
Selection / Ordering Information
Channel thresholds 2–100 µm, adjustable in 0.1 µm increments. Magnetic stirring, enclosed sampling chamber, automatic sampling-head lift, built-in printer and 5.7-inch touchscreen are documented. Reports can include contamination index and filtration efficiency. Operating temperature 10–40 °C. Confirm keyboard, calibration and delivery contents; qualification documents/services are separate paid options.
Downloads
Download the detailed product datasheet for full specifications and configuration information.
Download PDF — English · PDF